Advertisement

Do not pour fluid onto the Product
surface. Fluid seepage into the electrical
circuitry may cause the Product to fail.
Do not use spray cleaners on the
Product. Such action may force cleaning
fluid into the Product and damage
electronic components.
 Caution
To prevent possible damage to the Product
or to equipment under test, remove the input
signals before you clean the Product.

Outside

To clean the outside of the Product, disconnect from the
power supply and use only a damp cloth with mild
detergent.

Inside

It is possible that microbial growth can become present in
the transducers of the Product. It is recommended that
you clean the transducers at 3 month intervals. To clean
the inside of the Product, inject 20 ml of a warm water
and detergent solution into the Fluid Inlet Port. After 5
minutes, flush with clean water. Always pass water from
the fluid inlets to the outlets.

Test Fluid

The Product is intended to operate with de-ionized water
with added detergent. Fluids intended for use on patients,
such as high viscosity, oily, or corrosive substances will
damage the measurement system. Tap water can contain
contaminates which can also damage the transducer.
Test fluid can be made with de-ionized water and a
wetting agent such as MICRO-90. It is recommended that
a 0.1 % solution of MICRO-90 in de-ionized water
(preferably degassed) be prepared in volume for daily use
and kept in a sealed vessel. If the water makes too much
foam, then a 0.05 % dilution is recommended.
MICRO-90 is available from:
International Product Corp.
201 Connecticut Dr.
P.O. Box 70
Burlington, NJ 08016-0070 USA
Tel 609 386 8770
International Product Corp.
1 Church Row
Chistlehurst, Kent BR7 5PG United Kingdom
Tel. 0208 467 8944
Infusion Device Analyzer
Product Maintenance
11

Advertisement

Table of Contents
loading

Table of Contents