Setting up the Tracking System for eSieFusion Imaging
The tracking system consists of the following components:
electronics unit
transmitter
transducer sensors
needle tracking sensor
Connecting the Electronics Unit to the Ultrasound System
Note: The electronics unit must be installed by an authorized Siemens representative or by a
Siemens-approved third party.
The electronics unit must be connected to a USB port on the input/output panel of the
ultrasound system.
See also: Input/Output Panel Connections, System Setup, Chapter 3, Instructions for Use
Example of the trakSTAR2 electronics unit, back view.
1
RS-232 connection (not used)
2
Multi-unit synchronization connection (not used)
3
Power on/off switch
4
Power cord connection
5
Connection for analog switch (not used)
6
USB connection from ultrasound system
I n s t r u c t i o n s f o r U s e
9 e S i e F u s i o n I m a g i n g
9 - 2 5