Storing and Handling

The transesophageal transducer is a mechanically sensitive device. The distal tip is especially
sensitive to bumps, scrapes, and jarring. Take extreme care when handling and storing the
transducer. Never move the distal tip by hand. Always use the flex controls.
When you store the transesophageal transducer, ensure that the friction brakes are off, the
distal tip is straight, and the cables and flexible shaft are free of kinks.
Store the disinfected transducer in the original shipping case only after disinfecting. The
transesophageal transducer is fragile and should not be stored in the transducer holders on the
system.
Transporting the Transducer
Use the case for shipping or for moving the transducer from one site to another. Take care that
the distal tip is straight and that the cable is free of kinks as you carefully place the transducer
in the foam padding. Before you close the lid, ensure that nothing protrudes from the case.

Service and Repair

Do not attempt to repair or alter any part of the transesophageal transducer. Contact your local
Siemens service representative office immediately if your transducer appears to be damaged or
malfunctions in any way.