PERFORMANCE TEST
Test set-up:
Settings/procedure:
1 -
Apply a 1 kHz square-wave signal of 0.1V to input CH1. Output not terminated into 50Ω ('LZ'
must not appear in lower part of viewing area).
2 -
Press the AUTOSET key.
3 -
Set CH1 to 20 mV and DC coupled input.
4 -
Press the DISPLAY menu key.
5 -
Press X-DEFL softkey.
6 -
Select 'on' and 'ch1' from the X-DEFL menu.
7 -
Press the CH2 ON key and then the CH1 ON key; the result is that CH2 is on and CH1 is off.
Requirements:
-
Verify that two dots with a horizontal distance of 4.7 ... 5.3 divisions are displayed.

6.3.31 Horizontal deflection; X deflection coefficient via 'line'

The amplification of the horizontal amplifier via the line trigger signal is checked. Do this test only
when 220V line voltage is available.
Test equipment:
None
Settings/procedure:
1 -
Press the AUTOSET key.
2 -
Press the DISPLAY menu key.
3 -
Press X-DEFL softkey.
4 -
Select 'on' and 'line' from the X-DEFL menu.
Requirements:
-
Verify that a horizontal line of 4.3 to 7.7 divisions is displayed when the line voltage is 220V (rms).