PERFORMANCE TEST

6.3.41 Z-MOD sensitivity

This test checks the sensitivity of the Z modulation facility.
Test equipment:
Fluke 5500A mode: wavegen, wave square (Alternative: function generator PM 5136).
T-piece.
Test set-up:
Settings/procedure and requirements:
1 -
Apply a 1 kHz square-wave signal, duty cycle 50%, amplitude between 0 to +2.5V, to input
CH1. Adjust the generator to a signal of 2.5 Vpp with an offset of 1.25V.
2 -
Press the AUTOSET key.
3 -
Set MTB to 0.5 ms/div.
4 -
Set the trace of CH1 in mid position with the CH1 POS control.
5 -
Apply the same signal by means of the T-piece to the Z input (rear side).
6 -
Adjust TRACE INTENSITY so that only the bottom half of the squarewave is displayed. The
top half must be invisible (0.5 ms light on; 0.5 ms light off).
7 -
Decrease the input signal to 0.5 Vpp with an offset of 0.25V.
8 -
Set CH1 to 0.5V/division.
9 -
Verify that the top half of the square wave is visible at full intensity.
6 - 37