Horizontal Deflection; X Deflection - Fluke Autoranging CombiScope PM3370B Service Manual

Instrument
Hide thumbs Also See for Autoranging CombiScope PM3370B:
Table of Contents

Advertisement

6 - 20

6.3.23 Horizontal deflection; X deflection

The correct working of the X Y mode (X-DEFL 'on') is tested.
Test equipment:
Fluke 5500A mode: wavegen, wave sine (Alternative: function generator, PM5136)
Test set-up:
Settings/procedure:
1 -
Apply a 2 kHz sine-wave signal of 800 mV (pp) to input CH1.
2 -
Press the AUTOSET key and set CH1 to 0.1V/div.
3 -
Adjust the generator signal to a trace height of 8 divisions.
4 -
Press the DISPLAY menu key.
5 -
Press the X-DEFL softkey in the DISPLAY menu.
6 -
Select 'on' and 'ch1' as X-SOURCE in the X-DEFL menu.
7 -
Use the CH1 POS and X POS controls to obtain the display shown in the figure below.
Requirements:
-
Verify that a line with an angle of 45° is displayed.
Figure 6.4
MAT3837
X deflection
PERFORMANCE TEST