H5i technical specifications
Temperature
Physical
Material
Air tubing technical specifications
Air tubing
ClimateLine
ClimateLine
MAX
SlimLine
Standard
3m
Heated tubing temperature cut-out: ≤ 106°F (≤ 41°C)
Notes:
•
The manufacturer reserves the right to change these specifications without notice.
•
The temperature and relative humidity settings displayed for Climate Control are not measured
values.
•
Check with the service provider before using the SlimLine air tubing with devices other than the S9
or H5i.
•
The electrical connector end of the heated air tubing is only compatible with the H5i air outlet and
should not be fitted to the device or mask.
Humidifier performance
The following settings have been tested at 71.6°F (22°C) ambient temperature:
CPAP mask
pressure cm H
O
2
4
10
20
25
AH - Absolute Humidity in mg/L.
1
BTPS - Body Temperature Pressure Saturated.
2
30
Maximum heater plate temperature: 150°F (65°C)
Temperature cut-out: 165°F (74°C)
Maximum gas temperature at mask: ≤ 106°F (≤ 41°C)
Dimensions (L x W x H): 6.0" x 5.7" x 3.4" (153 mm x 145 mm x 86 mm)
Weight (standard water tub): Docking station and unfilled water tub 1.52 lb
(690 g)
Weight (cleanable water tub): Docking station and unfilled water tub 1.74 lb
(790 g)
Water capacity: To maximum fill line 380 mL
Docking station: Flame retardant engineering thermoplastic, aluminium
Cleanable water tub: Injection molded plastic, stainless steel and silicone seal
Standard water tub: Injection molded plastic, aluminium and elastomer seal
Length
Inner
diameter
6'6" (2.0 m)
0.6" (15 mm)
6'3" (1.9 m)
0.75" (19 mm)
6' (1.8 m)
0.6" (15 mm)
6'6" (2.0 m)
0.75" (19 mm)
9'10" (3.0 m)
0.75" (19 mm)
RH output %
Setting 3
90
95
95
100
Material
Flexible plastic and electrical components
Flexible plastic and electrical components
Flexible plastic
Flexible plastic
Flexible plastic
Nominal system output AH
Setting 6
Setting 3
100
100
100
100
Setting 6
10
11.5
11
12
13.5
, BTPS
1
2
18
21
18