6-1-1-3. Operation Check While the Machine Is Stopped

The test rod is not deformed.
Use a test rod with an appropriate diameter for inspection. A test rod is not included in the F3SG-R
product package.
The sensor must detect a test rod wherever it is in the detection zone. In other words, when a test rod
is inserted into the detection zone, the Stable-state indicator turns OFF and the ON/OFF indicator
turns red.
To check detection, move the test rod as shown in the diagram below.
When the external test function is used:
When 24 V Active is enabled, the ON/OFF indicator turns red when the TEST line is short-circuited
to 9 V to Vs.
When 0 V Active is enabled, the ON/OFF indicator turns red when the TEST line is short-circuited to
0 to 3 V.
When the external device monitoring function is used:
When the EDM line is open the ON/OFF indicator remains red regardless of the beam state of the
F3SG-R.

6-1-1-4. Checking that Hazardous Parts Stop While the Machine Operates

The hazardous parts stop immediately when a test rod is inserted into the detection zone at 3
positions: "directly in front of the emitter", "directly in front of the receiver", and "between the emitter
and receiver". (Use the appropriate test rod.)
The hazardous parts remain stopped as long as the test rod is in the detection zone.
The hazardous parts stop when the power of the F3SG-R is turned OFF.
The actual response time of the whole machine is equal to or less than the calculated value.
Start
Stop
Test rod
Checklists
F3SG-4RA-25-02TS
User's Manual
E
89