Figure 44
factory setting is shown
for autosampler
Table 53
Theory of Operation

Setting the 8-bit Configuration Switch

Setting the 8-bit Configuration Switch
The 8-bit configuration switch is located next to the GPIB connector.
Switch settings provide configuration parameters for GPIB address, serial
communication protocol and instrument specific initialization procedures.
8-bit Configuration Switch
1
8-bit Configuration Switch
Mode Select
1
GPIB
0
RS-232C
0
Reserved
1
TEST/BOOT 1
Switches 1 and 2 define which set of parameters (for example, for GPIB,
RS-232C, and so on) will be changed. Once the change has been completed,
the instrument must be powered up again in order to store the values in
the non-volatile memory.
In the non-volatile memory the parameters are kept, regardless of whether
you turn the instrument off and on again. They will be kept until the same
236
1
0
8
2
3
4
0
GPIB Address
1
Baudrate
0
Reserved
1
RSVD
SYS
5
6
7
Data
Parity
Bits
RSVD
RSVD
8
FC