Running Multiple-Vial Analyses
Recalibrating With Multiple Groups of Standards
C1 C2 C3 S13-S15 C1 C2 C3 S16-S18 C1 C2 C3 S19-S21 C1 C2 C3
• the calibration standards of type B are in positions 4, 5 and 6 of the autosampler
tray and the 9 sample vials are in positions 13 to 21, and
• the samples and calibration standards of type B use different methods.
The samples use METH2 and the calibration standards use METH3. These methods
contain the same analytical parameters and differ only in the analysis stoptime.
The example assumes that the sequence table is empty. If the sequence table is not
empty, use the All Lines item, accessed by pressing the Delete button [F6].
1 Select the Sequence button [F4] in the Analysis screen.
2 For sequence line 1, select the Insert button [F7] and enter:
Vial Range
#Inj.
Inj. Volume
Method
3 Select the Calibration button [F1] to display the Calibration Settings screen for
sequence line 1 and choose Edit.
4 Enter the following information:
Calibr. Vial range 1 to 3
#Inj.
Inj. Volume
Calibr. method
Recalibrate every 2 vials multi
Before
After
5 Select the Done button [F6] to accept entries.
6 Move the highlighted bar to line 2 (by pressing the Selection key down) and press the
Enter key.
7 For sequence line 2 enter:
Vial Range
#Inj.
124
7 to 12
1
Default
METH1
2
Default
METH1
On
Off
13 to 21
1