Technical Specifications
Dimensions (H x W x D)
5.6" x 9.8" x 11.5" (142 mm x 248 mm x 293 mm)
Weight
Flow generator with cover: 8.14 lb (3.7 kg)
Flow generator with Humidaire 2i (empty): 9.24 lb (4.2 kg)
Environmental conditions
Operating temperature: 41
Storage and transport temperature: -4
Humidity—operating, storage and transportation: 10%–95% non-condensing
Housing construction
Injection moulded plastic
Air tubing
1 x 6'6" (2 m)
Air filter
Synthetic fiber
Patient connection port
22 mm taper, compatible with EN 1281-1:1997 Anaesthetic & Respiratory
Equipment —Conical Connectors
Power supply
AC Input range: 110–120 V and 220–240 V; 50–60 Hz; 60 VA
DC Input range (if using DC converter): 12–13.8 V DC
Performance
Operating pressure range: 4–20 cm H
Sound pressure level
<30 dBA (tested in accordance with the requirements of ISO 17510-1)
Battery type
Lithium
Alarm specifications
•
High maximum mask pressure alarm
Stops air delivery at 25 cm H
•
High mean mask pressure alarm
Stops air delivery when mean mask pressure exceeds 15 cm H
over approximately 1 minute).
•
Low mask pressure alarm
Mask pressure is less than 3 cm H
patient is using the VPAP Adapt SV.
o
o
o
F to 104
F (+5
C to +40
o
o
F to +140
F (-20
O
2
O >700 ms.
2
O for longer than 5 seconds while the
2
o
C)
o
o
C to +60
C)
O (averaged
2
Technical Specifications
39