Users Manual
Table 7. Available Units of Measure (cont.)
Unit
Pressure
mbar
millibar = 0.001 bar
bar
14.7 psi
mmHg
Millimeters of mercury
inHg
Inches of mercury
cmH
0
Centimeters of water
2
inH
0
Inches of water
2
psi
Pounds per square inch
atm
Atmosphere
kPa
Kilopascals
32
Description

Special Tests Menu

Use the Special Tests menu for more tests.
Leak Tests calculate the lost volume over the
specified time. To do a leak test:
1. Select Menus > Special Tests.
2. Select the type of leak test:
Airway Pressure
High Pressure
Low Pressure
Ultra-Low Pressure
3. Use the keypad to enter the length of the
test.
4. Tap Compliance and use the keypad to
enter the compliance limit.
Use compliance to calculate volume lost. For
example, enter the ml of volume lost per
cmH
O. If blank, volume lost is not
2
calculated.
5. Tap Start.