Expiratory Test Connections

For expiratory connections, see Figure 7.
1. Connect the expiratory hose from the test
lung to the flow inlet on the Analyzer.
Inspiratory Hose
Test Lung
Expiratory Hose
Exhaust
Figure 7. Expiratory Breathing Circuit Connections
2. Use a standard breathing hose to connect
the ventilator to the exhaust port on the
Analyzer.
The Analyzer shows the flow of gas
delivered by the ventilator.