Troubleshooting - Hach HQ11d User Manual

Hqd series portable meter
Hide thumbs Also See for HQ11d:
Table of Contents

Advertisement

1. Pull the release tab on the battery cover and the remove the cover.
2. Remove the batteries.
3. Install 4 AA alkaline or 4 AA nickel metal hydride (NiMH) batteries.
Make sure that the batteries are installed in the correct polarity.
4. Replace the battery cover.
Figure 11 Battery replacement
1 Batteries
2 Release tab

Troubleshooting

Refer to the following table for common problem messages or
symptoms, possible causes and corrective actions.
Connect a Probe
Probe Not Supported Probe disconnected
Bootloader
X.X.XX.XX error
3 Battery cover
Error/Warning
Description
Probe disconnected
or connected
improperly
Software not
updated to most
current version
Problem with probe
or connected
improperly
Software not
updated to most
current version
Problem with probe
HQd meter does not
support IntelliCAL
probe
Software not
updated to most
current version.
Solution
Tighten the locking nut on the
probe connector.
Disconnect the probe and then
connect the probe again
To download the most current
version of the software, refer to
the applicable product page on
the manufacturer's website.
Connect a different IntelliCAL
probe to verify if problem is with
probe or meter
Tighten the locking nut on the
probe connector.
Disconnect the probe and then
connect the probe again.
To download the most current
version of the software, refer to
the applicable product page on
the manufacturer's website.
Connect a different IntelliCAL
probe to the meter to verify if
problem is with the meter or the
probe.
Contact Technical Support.
To download the most current
version of the software, refer to
the applicable product page on
the manufacturer's website.
English 25