Remove column headers

When transferred data is viewed in a spreadsheet program, the first row
of data contains headings to identify the type of data in each column. If
an application or post-processing method is used that is incompatible
with the headers, the column headers can be omitted.
1. Push the
key.
2. Select Column Headers.
3. Set the column headers to off.

Send data directly to a computer

Data can be transferred from any HQd series meter directly to a
computer when the HQ40d PC Application is installed. The data can be
sent in real time during data collection, or the entire data log can be
transferred.
To download the most current version of the software, refer to the
applicable product page on the manufacturer's website.
1. Install the HQ40d PC Application on the computer.
2. Turn off the meter. Make sure that the meter is connected to AC
power.
3. Connect the PC to the meter with a USB type B cable.
4. Turn on the meter.
5. Open the HQ40d PC Application on the computer. Click on the green
triangle in the menu bar to start a connection.
6. Collect the data in real time or transfer the data from the data log:
• Real time—when a data point is stored in the meter, the result is
sent simultaneously to the PC Application (refer to
measurement mode
on page 21).
• Data log—push the
key and select Send Data Log. Wait for
the display to show "Transfer Complete." The data is sent as a
comma separated values (.csv) file.
The data is shown in the HQ40d PC Application window.
18 English

Advanced operation

Security options

The Security Options menu is used to protect the meter setup and
method settings from unwanted changes. This menu is available in the
Full Access Options menu.
The Setup Measurement Mode, Date and Time, Temperature Units,
Language, Probe settings, Delete data log and Security Options screens
are disabled in the Operator Access Options menu. All menu options are
enabled in the Full Access Options menu.
Note:
The Full Access Options menu is shown when the key is pushed when
Security Options is OFF, whether or not a password has been set.
When the meter is powered on for the first time and Security Options is
selected, the display prompts the user to set a password. Until the meter
is shut off, pushing the
menu, even after Security Options is turned on and a password has
been set. After the meter is shut off and powered on again with Security
Options on, the Operator Access Options menu is displayed until a valid
password is entered.
Store the password in a safe and accessible place. If the specified
password is forgotten and Security Options is turned on, the operator is
locked out of the restricted menus. Contact technical support if the
password is lost.

Turn Security Options on

The Security Options and the Set Password options are used together to
prevent access to restricted menus.
Set the
key will still display the Full Access Options