Figure 13. LOI Menu
Switch HART revision mode
If the HART configuration tool is not capable of communicating with HART
Revision 7, the 2051 will load a Generic Menu with limited capability. The
following procedures will switch the HART revision mode from the Generic Menu:
1. Manual Setup > Device Information > Identification > Message
a. To change to HART Revision 5, Enter: "HART5" in the Message field
b. To change to HART Revision 7, Enter: "HART7" in the Message field
Step 6: Trim the transmitter
Devices are calibrated by the factory. Once installed, it is recommended to
perform a zero trim on gage and differential pressure transmitters to eliminate
error due to mounting position or static pressure effects. A zero trim can be
performed using either a Field Communicator or configuration buttons.
For instructions using AMS, please see the Rosemount 2051 Product Manual
(00809-0100-4107).
Note
When performing a zero trim, ensure that the equalization valve is open and all wet legs are filled
to the correct level.
It is not recommended to zero an absolute transmitter, model 2051TA.