September 2023
9
Zero trim the transmitter
Note
Emerson ships transmitters fully calibrated per request or by the
factory default of full scale (span = upper range limit).
A zero trim is a single-point adjustment used for compensating
mounting position and line pressure effects. When performing a zero
trim, ensure the equalizing valve is open and all wet legs are filled to
the correct level.
The transmitter will only allow three to five percent upper range limit
(URL) zero error to be trimmed. For greater zero errors, compensate
for the offset by using the XD_Scaling, Out_Scaling and Indirect
L_Type, which are part of the AI block.
Procedure
Zero trim the transmitter.
•
To use guided setup, navigate to Configure → Guided Setup and
select Zero Trim. The method will execute the zero trim.
•
To use manual setup, navigate to Overview → Calibration →
Sensor Trim and select Zero Trim. The method will execute the
zero trim.
Quick Start Guide
Quick Start Guide
29