May 2013

System Readiness

Confirm HART Revision Capability

If using HART based control or asset management systems, please confirm the
HART capability of those systems prior to transmitter installation. Not all
systems are capable of communicating with HART Revision 7 protocol. This
transmitter can be configured for either HART Revision 5 or 7.
For instructions on how to change the HART revision of your transmitter, see
.
page 15

Confirm Correct Device Driver

Verify the latest Device Driver (DD/DTM) is loaded on your systems to ensure
proper communications.
Download the latest Device Driver files at www.emersonprocess.com or
www.hartcomm.org.
Rosemount 2051 Device Revisions and Drivers
Verify the latest Device Driver (DD/DTM) is loaded on your systems to ensure
proper communications.
1. Download the latest DD at www.emersonprocess.com or
www.hartcomm.org.
2. In the Browse by Member dropdown menu, select Rosemount business unit of
Emerson Process Management.
3. Select desired Product
a. Within
numbers to find the correct Device Driver
Table 1. Rosemount 2051 Device Revisions and Files
Identify Device
Software
NAMUR
Release
Software
Date
Revision
April
1.0.0
2012
January
N/A
1998
1.NAMUR Software Revision is located on the hardware tag of the device. HART Software Revision can be read using a HART
capable configuration tool.
2. Device Driver file names us Device and DD Revision, e.g. 10_01. HART Protocol is designed to enable legacy device driver
revisions to continue to communicate with new HART devices. To access new functionality, the new Device Driver must
be downloaded. It is recommended to download new Device Driver files to ensure full functionality.
3.HART Revision 5 and 7 Selectable, Safety Certified. Local Operator Interface, Process Alerts, Scaled Variable, Configurable
Alarms, Expanded Engineering Units.
, use the HART Universal Revision and Device Revision
Table 1
Find Device Driver Files
HART
HART
Software
Universal
1
2
Revision
Revision
01
178
Device
2
Revision
7
10
9
5
5
3
Quick Start Guide
Review
Review Functionality
Instructions
Manual
Changes to
Document
Software
Number
See footnote 3 for
AA
list of changes.
3
N/A
3

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount 2051cf series

Table of Contents