Disassembly; Output Units - Omron E5AJ-A2HB Operation Manual

E5*j series
Table of Contents

Advertisement

Output Units

2-1

Disassembly

2-2

Output Units

Output Units
10
Before turning on the E5jJ, set its sensor type and control mode with the inter-
nal selectors. Refer to the following illustrations for disassembling the E5jJ to
access the internal switch settings.
After setting the internal switches, insert the internal mechanism into the case
until the front panel snaps with the hook.
Select the Output Unit according to the application and mount the Output Unit
into the socket on the E5jJ PCB of the as shown in the following illustration. The
E5CJ does not require an Output Unit. Refer to 1-3 Specifications for the output
ratings of the E5jJ.
The following Output Units are available.
Model
E53-R Relay Output Unit
E53-Q Voltage Output Unit
E53-Q3 Voltage Output Unit
E53-Q4 Voltage Output Unit
E53-C3 Linear Output Unit
E53-C3D Linear Output Unit
E53-V34 Linear Output Unit
E53-V35 Linear Output Unit
After mounting the Output Unit, be sure to secure it with the mounting bracket
provided with the Output Unit.
If the E53-C3, E53-C3D, E53-V34, and E53-V35 Linear Output Unit is used for
control output, no heater burnout alarm is available.
Each Voltage Output Unit is used for driving an SSR as shown in the following
illustrations.
Hook
Bracket
Output Unit
Socket
SPDT, 5 A at 250 VAC (resistive load)
NPN model, 40 mA at 12 VDC (with short-circuit
protecting circuit)
NPN model, 20 mA at 24 VDC (with short-circuit
protecting circuit)
PNP model, 20 mA at 24 VDC (with short-circuit
protecting circuit)
4 to 20 mA DC with a load of 600 Ω max.
0 to 20 mA DC with a load of 600 Ω max.
0 to 10 VDC with a load of 1 kΩ min.
0 to 5 VDC with a load of 1 kΩ min.
Section 2-2
Press the hook and
pull the front panel.