Fluke 190-104/190-204
Service Information
3. Press
After saving the calibration data, the calibration number and - date will be
updated if the calibration data have been changed and the data are valid.
The calibration number and - date will not change if:
- the calibration mode is entered and left without doing a calibration adjustment.
- only the contrast calibration adjustment (5.4) and/or the probe calibration is
done.
If you press
either calibrate the Test Tool again, or press
save and exit.
Possible error messages.
The following messages can be shown on the Test Tool display:
Proceed as follows:
The calibration date and number will not be updated. You must continue with the Final
Calibration!
5-18
Calibration data valid.
Save data and exit maintenance mode?
Calibration data valid indicates that the calibration adjustment procedure
is performed correctly. It does not necessarily mean that the Test Tool
meets the characteristics listed in Chapter 2.
F4
to save and exit.
YES
F3
NO
WARNING: Calibration data not valid.
Save data and exit maintenance mode?
If you did the WarmingUp and Pre-Calibration successfully (section 5.5), and you
want to store the Pre-Calibration data before continuing with the Final Calibration:
⇒ Press
F4
YES
When turning the Test Tool off and on again, it will show the message:
The instrument needs calibration.
Please contact your service center.
To return to the Maintenance mode, if you want to repeat the complete calibration:
⇒ Press
F3
NO
F1
Now press
calibrate the Test Tool, starting at section 5.5.
If you want to exit and maintain the old calibration data:
⇒ Turn the Test Tool off.
Note
Note 1
Note 2
,the Test Tool returns to the calibration mode. You can
F4
.
.
until the display shows
F4
,
EXIT
WarmingUp (CL 0200):IDLE
to
YES
, and