Fluke 190-104/190-204
Service Information
8. Next check channel B, C or D in succession:
9. When you are finished, set the 5500A to 0 (zero) Volt, and to Standby.
100 mV/div
200 mV/div
500 mV/div
4-10
Press
TRIGGER
and select B, C or D as trigger source with
Press B, C or D to assign vertical 'range' and 'move' to channel B, C or D,
Observe reading B, C or D.
Table 4-2. Volts DC Measurement Verification Points
Range
2 mV/div
5 mV/div
10 mV/div
20 mV/div
50 mV/div
1 V/div
2 V/div
5 V/div
10 V/div
20 V/div
50 V/div
5500A output V dc
+6.0 mV
-6.0 mV
+15.0 mV
-15.0 mV
+30.0 mV
-30.0 mV
+60.0 mV
-60.0 mV
+150 mV
-150 mV
+300 mV
-300 mV
+600 mV
-600 mV
+1.50 V
-1.50 V
+3.00 V
-3.00 V
+6.00 V
-6.00 V
+15.0 V
-15.0 V
+30.0 V
-30.0 V
+60.0 V
-60.0 V
+150 V
-150 V
F1
,
Input A, B, C, D Reading
+4.9 to +7.1
-4.9 to -7.1
+14.3 to +15.7
-14.3 to -15.7
+29.1 to +30.9
-29.1 to -30.9
+58.6 to +61.4
-58.6 to -61.4
+143 to +157
-143 to -157
+291 to +309
-291 to -309
+586 to +614
-586 to -614
+1.43 to +1.57
-1.43 to -1.57
+2.91 to +3.09
-2.91 to -3.09
+5.86 to +6.14
-5.86 to -6.14
+14.3 to +15.7
-14.3 to -15.7
+29.1 to +30.9
-29.1 to -30.9
+58.6 to +61.4
-58.6 to -61.4
+143 to +157
-143 to -157