Fluke 190-104/190-204
Service Information
8. When you are finished, set the 5500A to Standby.
UUT
Model
ALL
190-104
190-204
4-20
Table 4-8. Input A Trigger Sensitivity Test Points
UUT
Time base
Initial Input Voltage
200 ns/div
100 mV pp
10 ns/div
400 mV pp
10 ns/div
800 mV pp
10 ns/div
400 mV pp
10 ns/div
800 mV pp
5500A SC...
MODE levsin
Frequency
5 MHz
100 MHz
150 MHz
200 MHz
250 MHz
UUT
Trigger Amplitude
0.5 div
1 div
2 div
1 div
2 div