Input A Trigger Sensitivity Test - Fluke 190-104 Service Manual

Scopemeter 190 series ii
Hide thumbs Also See for 190-104:
Table of Contents

Advertisement

Using
(
800.0 μs
A
Using
7. Set the 5500A to source a 0.8 ms time marker (MODE marker).
8. Using
799.0 μs.
9. Examine the rising edge of the time marker pulse at the vertical height of the trigger
level indicator top. Verify that the rising edge is at the second grid line from the left.
The allowed deviation is ±3 pixels, see Figure 4-7.

4.6.9 Input A Trigger Sensitivity Test

Proceed as follows to test the Input A trigger sensitivity:
1. Connect the Test Tool to the 5500A as for the previous test (see Figure 4-6).
2. Select the following Test Tool setup:
Reset the Test Tool
Using
ranging, and lock the Input A sensitivity range on 2 V/div.
s
TIME
3. Using
Table 4-8.
4. Set the 5500A to source the leveled sine wave for the appropriate Test Tool model.
5. Adjust the 5500A output voltage until the displayed trace has the trigger amplitude
indicated under the last column of Table 4-8.
6. Verify that the signal is well triggered.
If it is not, press
manual Trigger Level adjustment. Adjust the trigger level and verify that the signal
will be triggered now. The trigger level is indicated by the trigger icon ( ).
7. Continue through the test points.
MOVE
move the trace to adjust the trigger delay time to 800.0 μs
).
s
TIME
ns
set the time base on 1 μs/div.
MOVE
move the trace to the right until the indicated trigger delay is
Figure 4-7. Time Base Verification
mV
RANGE
and
change the sensitivity range to select manual sensitivity
RANGE
V
ns
select the time base indicated under the second column of
TRIGGER
, then using
4.6 Scope Input A, B, C, D Tests
F3
enable the up/down arrow keys for
Performance Verification
190c-tb1.bmp
4
4-19

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

190-204190 ii series

Table of Contents