Fluke 190-104/190-204
Service Information
PM9093
2. Select the following Test Tool setup:
3. Press
4. Using
5. Set the 5500A to source a sine wave according to the first test point in Table 4-6.
6. Observe reading
7. Continue through the test points.
8. Next check channel B, C or D in succession:
9. When you are finished, set the 5500A to 0 (zero) Volt, and to Standby.
Model
all
4-16
PM9091
PM9092
A
B
C
D
Figure 4-5. 5500A Scope Output to Test Tool Input A, B, C, D
•
Recall the created setup (e.g. SETUP 1, see section 4.4.3): press
, select
RECALL
SCREEN+SETUP 1
•
Press
SCOPE
•
F1
Press
–
READINGS ON
•
F2
Press
–
READING ...
the arrow keys:
Reading 1, on A, Hz,
Reading 2, on B, Hz,
Reading 3, on C, Hz,
Reading 4, on D, Hz.
A
: using
s
TIME
ns
select the required time base setting.
As no 50Ω termination is applied, the 5500 leveled sine wave output amplitude will
be twice the set value.
and check to see if it is within the range shown under the
A
appropriate column.
Press
TRIGGER
and select B, C or D as trigger source with
Press B, C or D to assign vertical 'range' and 'move' to channel B, C or D,
Observe reading B, C or D
Table 4-6. Input A, B, C, D Frequency Measurement Accuracy Test
Time
5500A-SC... MODE
base
20 ms/div
wavegen, sine
CONNECT TO CHANNEL A, B, C, D IN PARALLEL
NO
50 OHM TERMINATION
NORMAL
SCOPE
FLUKE 5500A CALIBRATOR
, press
.
and select with
mV
RANGE
and
select range
RANGE
V
Voltage
Frequency
600 mVpp
SAVE
,
ENTER
to recall the setup .
F1
–
READINGS
50
mV/div for A.
F1
,
Input A&B Reading
16 Hz
15.90 to 16.10
F2
and with