Measuring Pulsed Signals - HP EPM-441A User Manual

Power meter
Hide thumbs Also See for EPM-441A:
Table of Contents

Advertisement

Power Meter Operation

Measuring Pulsed Signals

Measuring Pulsed Signals
The power meter can be used to measure the power of a pulsed signal. The
measurement result is a mathematical representation of the pulse power
rather than an actual measurement. The power meter measures the
average power of the pulsed input signal and then divides the
measurement result by the duty cycle value to obtain the pulse power
reading. The allowable range of values is 0.001 % to 100 %. The default
value is 1.000%.
If duty cycle is enabled then "Dty Cyc" is displayed.
Note
An example of a pulsed signal is shown in Figure 2-5.
2-28
Pulse measurements are not recommended using HP ECP-series
power sensors.
Figure 2-5: Pulsed Signal
Power
B
Duty Cycle = A
Time
A
HP EPM-441A User's Guide