HP EPM-441A User Manual page 117

Power meter
Hide thumbs Also See for EPM-441A:
Table of Contents

Advertisement

Edit Cal Table
Press this softkey to access a menu and screen which allows
you to modify, add and remove frequency and calibration
factors from the currently selected table and modify the table
name. Use the
between the name, frequency and calibration factors. Press
Change Insert
,
Change
Press this softkey to modify the highlighted parameter,
which is either a frequency, calibration factor or table name.
A pop up window displays the parameter. Use the
,
value. To confirm your choice press
Insert
Press this softkey to insert a new sensor calibration table
entry. You are prompted to enter the frequency and
calibration factor. The entry is inserted in ascending
frequency order.
Delete
Press this softkey to delete a sensor calibration table entry. If
you delete the frequency the corresponding calibration factor
is also removed and vice versa.
Done
Press this softkey to return to the measurement screen. The
softkeys from the first level of the "System/Inputs" menu are
displayed.
Table Off On
Press this softkey to enable and disable the highlighted sensor
calibration table for channel A. "ON" or "OFF" is displayed
beside each table to indicate its current state.
Done
Press this softkey to return to the measurement screen. The
softkeys from the first level of the "System/Inputs" menu are
displayed.
Input Settings
Press these softkeys to access a menu which allows you to modify
the settings of the averaging, duty cycle, range and offset of
channel A.
HP EPM-441A User's Guide
Front Panel Menu Reference "System/Inputs" Menu
,
,
and
or
Delete
as follows:
and
hardkeys to change the parameter
Menu Reference
hardkeys to move
,
.
Enter
3-17