No response from
instrument
Time for Periodic
Calibration. Please
contact the closest
Service Center.
Not Calibration data.
Set Calibration data.
Are you sure?
Check UV Settings.
UV adjustment not
possible.Change
Standard value or
tolerance settings,
or use a different
Fluorescent Standard.
UV adjustment not
possible. UV is set
to XX and cannot be
adjusted further in
the required direction.
Change Standard value
or tolerance settings,
or use a different
Fluorescent Standard.
Communication with
the instrument failed.
• The power of the
instrument is off.
• COM port setting
is incorrect.
• Communication
settings are incorrect.
Hardware initialization
at power on (of lens,
filter and trap positions)
is in progress.
It is time to perform
periodic calibration.
The calibration data for
while calibration is not found.
UV Settings has been
set to "100% Full".
Current settings did
not allow adjustment
of the filter position.
Current settings did
not allow adjustment
of the filter position.
Turn on the power of the
instrument.
Check and correct
the COM port and
communication settings.
After turning the power on,
wait for ten and several
seconds until the lens, trap
and UV filter positions are
initialized.
Contact a KONICA
MINOLTA-authorized
service facility and request
periodic calibration.
Load the calibration data
into the instrument.
Set the UV Settings to
"UV adjustment" .
Change Standard value
or tolerance settings,
or use a different
Fluorescent Standard;
then retry "UV adjustment".
Change Standard value
or tolerance settings,
or use a different
Fluorescent Standard;
then retry "UV
adjustment".
30
15
-
-
20
Refer to the
SpectraMagic™
NX (optional)
instruction manual.
34
34