Option 1D5, High Stability Frequency Reference
This option offers ±0.05 ppm temperature stability from 0 to 60 °C (referenced to 25 °C).
Option 002, Harmonic Mode
This option provides measurement of second or third harmonics of the test device's
fundamental output signal. Frequency and power sweep are supported in this mode.
Harmonic frequencies can be measured up to the maximum frequency of the receiver.
However, the fundamental frequency may not be lower than 16 MHz.
Option 004, Step Attenuator (8753ET only)
This option adds a 55 dB step attenuator into the RF output path.
Option 006, 6 GHz Operation
This option extends the maximum source and receiver frequency of the analyzer to 6 GHz.
Option 010, Time Domain
This option displays the time domain response of a network by computing the inverse
Fourier transform of the frequency domain response. It shows the response of a test device
as a function of time or distance. Displaying the reflection coefficient of a network versus
time determines the magnitude and location of each discontinuity. Displaying the
transmission coefficient of a network versus time determines the characteristics of
individual transmission paths. Time domain operation retains all accuracy inherent with
the correction that is active in of such devices as SAW filters, SAW delay lines, RF cables,
and RF antennas.
Option 011, Receiver Configuration (8753ES only)
This option allows front panel access to the R, A, and B samplers and receivers. The
transfer switch, couplers, and bias tees have been removed. Therefore, external accessories
are required to make most measurements.
Option 014, Configurable Test Set (8753ES only)
This option provides direct access to the internal components of the analyzer. Front panel
access is provided to the A and B sampler ports, the COUPLER/SWITCH ports, the RF IN
and RF OUT ports, and the R CHANNEL ports. This option offers improved sensitivity
and the ability to add peripheral devices in multiple configurations.