Switching a Relay Channel

Switching a Relay Channel
Switch modules can be used to route signals to and from your test
system. This is achieved by closing or opening the relay channels on
these modules.
From the front-panel, you can open or close one relay channel at a
time. However, over the remote interface, multiple relay channels can
be operated by a single command if a channel list is specified. In
addition, these open or closed states can be stored, and a stored
channel setup can be included in a scan list (see "State Storage" on
page 106).
Whenever a switch module is reset, all of the closed relay channels on
the module will be opened.
Whenever the instrument is turned on or reset, all of the closed relay
channels in the instrument will be opened.
Front-Panel Operation: Select a channel, press Open or Close.
Select a slot, press and hold Card Reset to open all channels on the
selected module.
Press Shift, then press and hold Card Reset to open all channels in
the instrument.
Remote Interface Operation: The following commands open and close
multiple channels on the modules in slots 1 and 2.
OPEN (@101,103:107,207)
CLOS (@101,103:107,207)
SYST:CPON 1
When an Agilent N2260A or N2266A is configured to the 1-wire mode,
only one channel on the module can be closed at a time.
82
! Open multiple channels.
! Close multiple channels.
! Open all channels on the
module in slot 1.