Instrument Errors - Agilent Technologies Agilent 3499A User Manual

Switch/control system
Hide thumbs Also See for Agilent 3499A:
Table of Contents

Advertisement

-420
-430
-440
100
101
6
102
103
104
110
111
Chapter 6 Error Messages

Instrument Errors

Query UNTERMINATED
The instrument was addressed to talk (i.e., send data over the interface)
but a command has not been received which sends data to the output
buffer. For example, you may have executed a ROUTe command (which
does not generate data) and then attempted to read data from the remote
interface.
Query DEADLOCKED
A command was received which generates too much data to fit in the
output buffer and the input buffer is also full. Command execution
continues but all data is lost.
Query UNTERMINATED after indefinite response
The *IDN? command must be the last query command within a command
string. The *IDN? command returns an indefinite length string which
cannot be combined with any other query command.
Example: *IDN?;*STB?
Instrument Errors
Number of SAV/RCL out of range
This error will occur if a number included in *SAV or *RCL is out of range.
Up to 50 instrument setups can be stored using Firmware REV 4.0
(numbered 1 to 50). Up to 10 setups can be stored using Firmware REV
1.0/2.0/3.0 (numbered 1 to 10).
Examples: *SAV 52 or *RCL 0
Unable to recall - scan is running
Unable to recall - memory is empty
Unable to recall - modules were changed
Before recalling a stored channel setup, the instrument verifies the same
module types are installed in each slot. This error indicates that the
instrument has detected one or more modules have been replaced with
other module types or have been removed from the instrument.
Unable to store - scan is running
Slot number out of range
The specified slot number is invalid. The channel number has the form
(@snn), where s is the slot number and nn is the channel number.
Example: OPEN (@604)
Data out of range
The data for some commands is invalid.
Example: SOUR:DIG:DATA:BYTE:VAL 266 (valid data should be 0-255)