I/O LINES
Bit Number:
Maximum Voltage:
Maximum Sink Current:
Output Characteristics:
Input Characteristics:
HANDSHAKE LINES
Maximum Voltage:
Output Characteristics:
Input Characteristics:
9
Chapter 9 Specifications

N2264A Multifunction Module

N2264A Multifunction Module
The maximum number of channels that can be closed simultaneously is
10 (including the GP and High-Current GP relays). The maximum
carrying current is 15 A (including the GP, High-Current GP and
Digital I/O.
16-bit Digital I/O Function
16
Line-Chassis:
Per Bit:
Vout (high):
Vout (low):
Vin (high):
Vin (low):
Line-Chassis:
Vout (high):
Vout (low):
Iout (low):
Vin (high):
Vin (low):
310
+ 42 V dc
600 mA
≥ 2.4 V @ I ≤ 10 mA output
≤ 0.8 V @ I ≤ 600 mA input
≥ 2.0 V
≤ 0.8 V
+5 V dc
≥ 2.4 V @ I ≤ 400 µA output
≤ 0.5 V @ I ≤ 1 mA input
< 25 mA (when shorted to +5 V)
≥ 2.0 V
≤ 0.8 V