Display Control

Display Control
You can turn off the 3499A/B/C display (for security or increased
processing speed for example). You can also write a message of up to 13
characters to the front-panel display.
The display cannot be turned off from the front-panel.
When the display is turned off, the entire front-panel display goes
blank except for the ADRS and RMT annunciators (the ERROR
annunciator will also be on if there are errors), and all keys except
Local are locked.
If the display is turned off, pressing Local causes the instrument to
return to local operation, the display will be turned back on.
The display is automatically turned on when power is cycled, or after
a *RST command.
When the display is on, you can send a message (up to 13 characters)
to display on the front-panel from the remote interface. If you attempt
to send more than 13 characters, only the first 13 characters can be
displayed. The characters can be letters (A-Z), numbers (0-9), and
some special characters (" ' space ( ) * + - , . : ; / \).
Remote Interface Operation:
DIAGnostic:DISPlay:STATe OFF
DIAGnostic:DISPlay:STATe ON
DIAGnostic:DISPlay "Scan finished"
110
! Turn off the display.
! Turn on the display.
! Display the message on the
front-panel.