Output Current

The Calibrator is a fully programmable precision source of DC current from 0 mA
to 100 mA. The Calibrator can only output positive (+) values.
To output current
1. Connect the Calibrator to the UUT as shown in Figure 2-5.
2. If necessary, press B to select mA.
3. Use the numeric keypad to enter the desired output value and press J. You
can also adjust the output value by pressing M O. Press L N to select a
different digit to edit.
4. Press A activate output.
If OVER LOAD appears on the display you may not have a
completed circuit loop. When the circuit is complete, the message
will disappear.
725
MULTIFUNCTION CALIBRATOR
V mA
ZERO
LOOP
V mA
Hz
MEAS
SOURCE
TC
RTD
˚
C
˚
F
STORE
100%
SETUP
RECALL
25%
25%
0%
525A
TEMPERATURE / PRESSURE CALIBRATOR
mA
VOLTS
100mA MAX
HI
HI
OUTPUT
100V MAX
LO
LO
20V PK
MAX
4W RTD
HI
INPUT
SENSE
LO
20V PK
MAX
Figure 2-5. Connection for Setting Current Output