Output DC Voltage

The Calibrator is a fully programmable precision source of DC voltage from 0 V
to 100 V. The Calibrator can only output positive (+) values.
To output DC voltage
1. Connect the Calibrator to the UUT as shown in Figure 2-3.
2. Press B for V out display.
3. Use the numeric keypad to enter the desired output value and press J. You
can also adjust the output value by pressing M O. Press L N to select a
different digit to modify.
4. Press A to activate output.
For safety purposes, the Calibrator resets to standby when output is
set to 30 V or greater.