Dc Current Specifications, Output; Resistance Specifications, Output; Resistance Specifications, Input - Fluke 525A User Manual

Temperature/pressure calibrator
Hide thumbs Also See for 525A:
Table of Contents

Advertisement

DC Current Specifications, Output

1
Ranges
0 to 100.000 mA
1. All outputs are positive only.
Ranges
0 to 100.000 mA

Resistance Specifications, Output

1
Ranges
5 to 400.00 Ω
5 to 4.0000 kΩ
1.
Continuously variable from 0 to 4 kΩ.
2.
For currents lower than shown, the floor adder increases by Floor
I
min
/I
actual
.
For example, a 500 µA stimulus measuring 100Ω has a floor uncertainty of 0.025Ω x 1 mA/500
µA = 0.05 Ω.

Resistance Specifications, Input

Ranges
0 to 400.00 Ω
401 to 4001.00 Ω
Absolute Uncertainty,
tcal ±5 °C
± (ppm of output +µA)
90 days
1 year
85
2
100
Bandwidth
0.1 to 10 Hz p-p
2000 nA
Absolute Uncertainty,
tcal ±5°C, ± Ω)
90 days
1 year
0.025
0.03
0.25
0.3
Absolute Uncertainty,
tcal ±5°C, ±(ppm of output +Ω)
90 days
35
0.003
40
35
0.03
40
DC Current Specifications, Output
Compliance
Resolution
1 µA
2
Noise
10 Hz to 10 kHz rms
Resolution
0.01 Ω
0.1 Ω
(new)
1 year
Resolution
0.001 Ω
0.003
0.01 Ω
0.03
Specifications
Maximum
Maximum
Inductive
Voltage
Load
100 µH
10 V
Bandwidth
20 µA
Allowable Current
1 to 10 mA
250 µA to 1 mA
= Floor
(old)
x
Stimulus Current
1 mA
0.1 mA
7
1
7-3

Advertisement

Table of Contents
loading

Table of Contents