Acyclic Extended Diagnostics (General Fault Codes) - Siemens SITRANS LR200 Instruction Manual

Hide thumbs Also See for SITRANS LR200:
Table of Contents

Advertisement

Acyclic Extended Diagnostics (General Fault Codes)

In addition to the extended diagnostics available by cyclic data exchange (shown above),
further extended diagnostics are available via acyclic communications. This consists of
six bytes. Please see
for the location of the "Extended Diagnostics".
Note:
Certain fault codes (identified by an asterisk [*] in the table below) will persist
until a manual reset has been performed (see
LCD
Meaning
display
The device was unable
to get a measurement
S:0
for the LOE timer period.
Internal electronics
S:2
failure.
The Input parameters
High Calibration Point
S:10
and Low Calibration
Point are equal.
Internal temperature
S:11
sensor failure.
Internal temperature of
the device has been
exceeded: it is operating
S:12
*
outside its temperature
range.
7ML19985HP01
SITRANS LR 200 (PROFIBUS PA) – INSTRUCTION MANUAL
Appendix J: Asynchronous Communications Data Map
Corrective Action
Check the application and echo
profile to determine optimum
configuration for the application,
to ensure the device can obtain a
measurement.
Cycle the power. If the condition
persists, call the factory.
Check the calibration settings of
the device. Ensure that the
settings for High Calibration Point
and Low Calibration Point are
different.
Return the device to the factory.
Lower the ambient temperature
enough to cool the device.
This fault code will persist until a
manual reset is performed using
PDM or the LCD interface.
Fault Reset
on page 20).
on page 75,
Byte Bit
0
0
2
3
4
1
5
Page 29