General Fault Codes

Notes:
• If more than one fault is present, the device status indicator and text for each fault
alternate at 2 second intervals.
• Some faults cause the device to go to Failsafe mode (Fault 52). These are indicated with an
asterisk (*).
Code /
Icon
S: 0
*
The device was unable to get a mea-
surement within the Failsafe Timer
period. Possible causes: faulty
installation, antenna buildup, foam-
ing/other adverse process condi-
tions, invalid calibration range.
S: 2
*
The device is operating in a low
power condition that is outside its
operating range. As a result, a valid
measurement has not been taken
for the failsafe timer period, and the
device will be put into failsafe mode.
Page 94
SITRANS LR250 (HART) – INSTRUCTION MANUAL
• Data exchanged
• No data exchange
• Write access enabled
• Write access disabled
General Fault Codes
Meaning
Meaning (Continued)
Corrective Action
Ensure installation details are cor-
rect. Ensure no antenna buildup.
Adjust process conditions to mini-
mize foam or other adverse condi-
tions. Correct range calibration. If
fault persists, contact your local Sie-
mens representative.
Correct the power supply (resis-
tance or voltage).
7ML19985JE01