716 Calibration

mA Measure

1. Hold down the U and the D keys at the same time, then press the green
2. Connect the test leads from the AUX jacks of the Fluke 5520A to the mA jacks on
3. Set the Fluke 5520A to 0.00000 mA and press the O key.
4. Press any key on the 716 Calibrator. The display should change to:
5. Set the Fluke 5520A to 24.0000 mA.
6. Press any key on the 716 Calibrator. Display should change to:
7. Turn the green
717 Calibration (Earlier than V2.0)
mA Measure
1. Hold down the U and the D keys at the same time, then press the green
2. Connect the test leads from the AUX jacks of the Fluke 5520A to the mA jacks on
3. Set the Fluke 5520A to 0.00000 mA and press O.
4. Press any key on the 717 Calibrator. The display should change to:
5. Set the Fluke 5520A to 24.0000 mA.
6. Press any key on the 717 Calibrator. Display should change from:

Pressure Measure

1. Remove test leads from the 717 Calibrator and carefully attach the pressure fitting to
64
When the display shows CAL momentarily, release the U and the D keys
immediately. After CAL shows, the display will automatically read:
CAL
0.000mA
the 716 Calibrator (Black to Black and Red to Red).
-------
0.000mA
then to:
CAL
24.000mA
-------
24.000mA
The unit will reset itself. All segments will be displayed momentarily and change to:
-------
mA
O
power key OFF and push the Y key on the Fluke 5520A.
When the display shows CAL momentarily, release the U and the D keys
immediately. After CAL shows, the display will automatically read:
CAL
0.000mA
the 717 Calibrator (Black to Black and Red to Red).
-------
0.000mA
then to:
CAL
24.000mA
-------
24.000mA
to:
rAnGE
the pressure jack on the 717 Calibrator.
O
key.
O
key.