71X Series
Calibration Manual
Input pressure
0 psi
6 psi
12 psi
18 psi
24 psi
30 psi
24 psi
18 psi
12 psi
6 psi
-12 psi *
* Vacuum is only specified for V1.3 and later. To display the firmware version, start with the 713 off, press and
hold M, then press the power button.

mA Measure Verification

The display should indicate:
1. Connect a test lead from Red AUX terminal of the 5520A to the Fluke 8508A I+
2. Connect a test lead from Black AUX terminal of the 5520A to the 713 Calibrator
3. Connect a test lead from I- terminal of the Fluke 8508A to the 713 Calibrator mA
4. Adjust the 5520A if necessary so that the current shown on the Fluke 8508A is the
5. Verify that the display readings on the 713 calibrator are within the limits.
6. Disconnect the test leads and turn off the 713 Calibrator.
32
Table 43. 713 Pressure Verification
713 30G
Display Reading
-0.015 to 0.015
5.985 to 6.015
11.985 tp 12.015
17.985 to 18.015
23.985 to 24.015
29.985 to 30.015
23.985 to 24.015
17.985 to 18.015
11.985 tp 12.015
5.985 to 6.015
-12.015 to -11.985
0.000mA
0.00 PSI
terminal.
Com terminal.
terminal.
same as the 5520A values shown in Table 44.
Table 44. 713 mA Measure Verification
Fluke 5520A
4.0000 mA
12.0000 mA
24.0000 mA
713 100G
Input pressure
0 psi
20 psi
40 psi
60 psi
80 psi
100 psi
80 psi
60 psi
40 psi
20 psi
-12 psi*
Fluke 713
3.998 mA to 4.002 mA
11.996 mA to 12.004 mA
23.993 mA to 24.007 mA
Display Reading
-0.05 to 0.05
19.95 to 20.05
39.95 to 40.05
59.95 to 60.05
79.95 to 80.05
99.95 to 100.05
79.95 to 80.05
59.95 to 60.05
39.95 to 40.05
19.95 to 20.05
-12.05 to -11.95