Firmware Earlier Than V2.0 - Fluke 712 Calibration Manual

71x series process calibrators
Hide thumbs Also See for 712:
Table of Contents

Advertisement

71X Series
Calibration Manual
Displayed
Pressure Units
psi
O at 4 °C
inH
2
O at 20 °C
inH
2
O at 4 °C,
cmH
2
O at 20 °C
cmH
2
bar
mbar
kPa
inHg
mmHg
2
kg/cm
Range
24 mA
Overload protection: 125 mA, 250 V fast acting fuse
Temperature coefficient: 0.005 % of range per °C for temperature ranges – 10 °C to 18 °C and 28 °C to
55 °C
714 Specifications
714 Calibrator specifications vary based on the version of the instrument. To display the
firmware version for your instrument, start with the 714 off, press and hold {, then
press the power button. Find the section heading below for the displayed version and use
the specification tables in that section to test and calibrate the instrument.

Firmware Earlier than v2.0

TC Type
J, K, T, E, L, U
B, R, S
Maximum input voltage: 30 V
Temperature Coefficient: 0.05 x specified accuracy per °C for temperature ranges – 10 °C to 18 °C and
28 °C to 55 °C
12
Table 13. 713 Pressure Range and Resolution
Model 713 30G
Range and Resolution
30.000 psi
830.4 inH
831.9 inH
2109.0 cmH
2113.0 cmH
2.0685 bar
2068.5 mbar
206.85 kPa
61.080 inHg
1551.3 mmHg
2.1090 kg/cm
Table 14. 713 30G and 713 100G DC mA Input
Resolution
0.001 mA
Table 15. 714 Temperature Measure and Thermocouple Simulate
Resolution
0.1 °C
1 °C
O
2
O
2
O
2
O
2
2
±(% of Reading + Counts)
Error
± (0.3 °C + 10 µV)
± (0.3 °C + 10 µV)
Model 713 100G
Range and Resolution
100.00 psi
2768.0 inH
O
2
2772.9 inH
O
2
7030 cmH
O
2
7043 cmH
O
2
6.895 bar
6895 mbar
689.5 kPa
203.6 inHg
5171 mmHg
2
7.030 kg/cm
Accuracy,
0.025 + 1
Reference Junction Error
± 0.2 °C
± 0.2 °C