Firmware V2.0 And Later - Fluke 712 Calibration Manual

71x series process calibrators
Hide thumbs Also See for 712:
Table of Contents

Advertisement

Range
-10 mV to 75 mV
Maximum input voltage: 30 V

Firmware V2.0 and Later

Table 17. 714 Temperature Measure and Thermocouple Simulate
TC Type
J
K
T
E
R
S
B
L
U
Maximum input voltage: 30 V
* Includes 0.2 °C cold junction compensation (CJC) error
Range
-10 mV to 75 mV
Maximum input voltage: 30 V
Maximum source current is 1.0 mA
Table 16. 714 Millivolt Measure and Source
Resolution
0.01 mV
Range °C
-210.0 to 0.0
0.0 to 800.0
800.0 to 1200.0
-200.0 to 0.0
0.0 to 1000.0
1000.0 to 1372.0
-250.0 to 0.0
0.0 to 400.0
-250.0 to -100.0
-100.0 to 1000.0
-20.0 to 0.0
0.0 to 1787.0
-20.0 to 0.0
0.0 to 1767.0
600.0 to 800.0
800.0 to 1000.0
1000.0 to 1820.0
-200.0 to 0.0
0.0 to 900.0
-200.0 to 0.0
0.0 to 600.0
Table 18. 714 Millivolt Measure and Source
Resolution
0.001 mV
Accuracy
± (0.025 % of range (75 mV) + 1 count)
Accuracy °C *
0.6
0.4
0.5
0.8
0.5
0.7
0.8
0.4
0.8
0.4
2.0
1.4
2.0
1.4
1.4
1.5
1.7
0.45
0.4
0.7
0.45
Accuracy
0.015 % ± 10 µV
Process Calibrators
Specifications
13