Technical Information
Planar D6015TM Series Medical Monitor Specifications
Display Panel
Display Colors
Dimensions:
Weight:
Display Area
Response:
Viewing Angle:
Contrast Ratio
Brightness:
Pixel Pitch
Reliability
Video Interface
Scanning Frequency
Factory Preset Modes
Power Management
Power Consumption
Net Weight
Power Supply:
Options
Temperature: Operating
Humidity:
Altitude
Shock
Audio
Regulatory
18
15-inch (381 mm) XGA active matrix LCD (AMLCD)
262,144
w/bracket
408 x 333 x 85 mm (W x H x D)
w/desk stand
408 x 387 x 175 mm (W x H x D)
w/bracket
4.9 kg (10.8 lbs.)
w/desk stand
6.9 kg (15.2 lbs.)
304.1 x 228.1 mm (W x H), 15inch (381 mm) diagonal
Wht>Blk
10 ms typical
Blk>Wht
15 ms typical
H or V
80 ° typ., contrast ratio >10
300:1 typical
Non-touch
228 cd/m
w/SAW
225 cd/m
w/RT
200 cd/m
0.297 x 0.297 mm
40k hours MTBF, excluding backlight
VGA-compatible analog RGB (15-pin D-Sub), Separate
Sync. / Composite Sync.
24-61 kHz (horizontal), 50-90 Hz (vertical)
25
Meets VESA DPMS, EPA
30 W on, 4.5 W standby (not including power adapter)
6.9 kg, 15.2 lb
12V / 3A, 36W, Input power 100 Vac to 240 Vac at 50 to
60 Hz, AC Adapter (external)
Desk Stand, Resistive Touchscreen, SAW Touchscreen
0 to 40 °C
Non-operating
-20 to 65 °C
Operating
10 to 90% RH non-condensing, per IEC 68-2-3
Non-operating
0 to 95% RH non-condensing, per IEC 68-2-30
0 to 10k ft Operating, per IEC 68-2-13, 4 hr
0 to 40k ft. Non-operating, per IEC 68-2-13, 4 hr
50g, 11ms duration. Operating/Non-operating. Half sine
with 3 shocks on each of six axes per IEC 68-2-27.
Two 1-Watt speakers with amplifier
TÜV Safety Mark, CSA (C US Mark), CE Mark-MDD
2
typical
2
typical
2
typical
Operations Manual (OM700-02)