External Calibration (P6 Zcl); Restore Default Setting (P7 Rst) - Excell QWS User Manual

Waterproof indicator
Table of Contents

Advertisement

EXCELL PRECISION CO., LTD.

3-5 External calibration (P6 ZCL)

 This function is not available for approval models.
Enter external calibration"P6 ZCL" and press
Press
key to exit without saving.
.

3-6 Restore Default Setting (P7 rSt)

 This function is not available for approval models.
Enter restore default setting "P7 rSt" and press
Restart the scale
automatically
 The defaults are included the following:
1) External weight calibration
2) HI/LO/OK setting values
3) Parameters in P1 rEF
4) Sampling settings for the counting function
5) Backlight and autopower off setting
02027011
key to confirm and enter sub-menu or
Display last setting. Press
value. Press
kg
Press
key to save the setting or press
key to exit without saving and back to
last step.
0
.
1
key to restart the scale.
23
key to adjust
or
key to move cursor
Restart and back to
kg
weighing mode
ZSME300000342