Calibrating Using The Ir Link - Honeywell GasAlert Max XT II User Manual

Hazardous gas detection system
Table of Contents

Advertisement

GasAlertMaxXT II
User Manual
3. To ensure the readings are accurate, apply the verifica-
tion gas for the same amount of time as was applied to
the sensor when it was calibrated.
Example: H
S span time 2 minutes therefore, apply
2
verification gas for 2 minutes.

Calibrating Using the IR Link

If the Cal IR option is enabled, the sensors must be calibrated using
either the IR Link or MicroDock II station.
48
To calibrate using the IR Link, complete the following:
1. From the PC, open Fleet Manager II.
2. Click Administration.
3. From the Administration toolbar, click Login / Logout
to access the Enter Password dialog box. Enter Admin
and click OK.
4. From the Devices toolbar, click Detector Configuration.
5. Activate the detector.
Calibration can be performed during the startup tests
when CAL DUE NOW displays, or when the detector in
normal operation.
6. From Fleet Manager II, click the Calibrate button located
at the bottom of the window.
The Calibrate Device dialog box displays.
Figure 7. Calibrate Device Dialog Box