Connect With Intellidox; Connect With An Ir Link - Honeywell BW Ultra User Manual

Portable five-gas detector with internal pump
Hide thumbs Also See for BW Ultra:
Table of Contents

Advertisement

General Sensor Settings (for all sensors)
Sensor Enable/Disable
l
Calibration Gas Concentration
l
Note: When the operator configures the Honeywell BW Ultra using Safety Suite Device
Configurator (SSDC), Honeywell strongly recommends reviewing the detector's settings before
the operation to ensure that settings were applied successfully and comply with performance
requirements.
Custom configuration created in Safety Suite Device Configurator (SSDC) can be used to
configure detector settings.
Example: Five detectors must have the same bump test reminders and alarm setpoints. Each
detector could be configured separately or Safety Suite Device Configurator (SSDC) can be
utilized to create a custom settings configuration. This configuration can then be loaded on
each detector. This saves time and allows settings to be managed from one location.
Connect with IntelliDoX®
If a detector's calibration is overdue and the forced calibration feature is enabled, the
calibration can be performed with an IntelliDoX docking station or through the calibration
option from the detector's main menu.

Connect with an IR Link

The detector can be paired with an IR link, aka dongle. There is an IR connection at the bottom
of the detector that allows Safety Suite Device Configurator (SSDC) configurations to be
efficiently transferred to multiple detectors. The IR link will also allow new firmware to be
transferred to detectors or data/event logs to be transferred to Safety Suite Device Configurator
(SSDC).
Note: You should have the IR Connectivity Kit (sold separately) to transfer the data from a
computer to the detector.