Safe Mode; Confidence Beep; Latching Alarms; Force Calibration - Honeywell GasAlert Max XT II User Manual

Hazardous gas detection system
Table of Contents

Advertisement

GasAlertMaxXT II
User Manual

Safe Mode

If enabled, SAFE displays continuously on the LCD unless an alarm
condition occurs.
If an alarm condition occurs, the LCD displays the real-time readings for
each sensor.

Confidence Beep

If enabled, the confidence beep provides continuous confirmation that
the detector is operating correctly. To define how often the detector
beeps (every 1-120 seconds), enter the value in the
val
field.
Confidence beep automatically disables during a low battery
alarm.
The detector is shipped with the Confidence Beep option disabled.

Latching Alarms

If enabled, a detector alarm persists until the alarm is acknowledged and
gas concentrations are below the low alarm setpoint. The audible alarm
can be temporarily deactivated by pressing C, but the LCD continues to
24
Confidence Inter-
Note
display the high peak concentration until the alarm condition no longer
exists.
The detector is shipped with the Latching Alarms option disabled.

Force Calibration

Force Calibration Enabled: If enabled and a sensor(s) is past due for
calibration, the following screen displays.
The sensor(s) must be calibrated to continue and enter normal
operation. For complete instructions, refer to Calibration.
If the calibration is unsuccessful for any of the sensors, the
detector deactivates.
Force Calibration Disabled: If disabled, the CAL DUE NOW screen
displays. Press C to acknowledge the warning and enter normal
operation.
The detector is shipped with the Force Calibration option disabled.
If overdue for calibration, BW recommends the sensor(s) be
calibrated immediately.
Note
Note

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents