Pump Test on Startup

The diffusion cover must be attached to the detector to activate
the pump and initiate the pump test.
The maximum hose length for sampling is 75 ft (22m).
5. In cold temperatures, the pump may require a short
period of time to warm up before operating. If this is
required, the following screen displays.
The LCD displays a countdown of the time remaining
(in seconds) for the pump to warm up.
If the Force Block Test option is enabled, the detector
performs a pump test. The following screen displays.
Note
a Caution
Using your finger, block the end of the hose. The follow-
ing screen displays.
If the hose is not blocked and unblocked within 2.5 minutes,
the detector will assume the pump has failed and deactivates.
Successful Pump Test: If the pump test is successful,
the following screen displays.
GasAlertMaxXT II
Startup Tests
Note
13