Alarm And Saturation Level Configuration; Alarm And Saturation Levels For Burst Mode - Emerson Rosemount 3051S Series Reference Manual

Scalable pressure, flow, and level solution with hart protocol
Hide thumbs Also See for Rosemount 3051S Series:
Table of Contents

Advertisement

Reference Manual
00809-0100-4801, Rev EA
August 2010
Alarm and Saturation
Level Configuration
1, 4, 2, 7
Traditional Fast Keys
2, 2, 1, 7
Device Dashboard
Fast Keys
Alarm and Saturation
Levels for Burst Mode
To configure alarm and saturation levels with a Field Communicator or AMS
perform the following procedure:
Field Communicator v3.3
1. From the HOME screen, follow the fast key sequence.
2. Select 7, Config. Alarm Level to configure alarm levels.
3. Select OK after setting the control loop to manual.
4. Select OK to acknowledge current settings.
5. Select desired setting, if "OTHER" is selected enter HI and LO custom
values.
6. Select OK to acknowledge the loop can be returned to automatic control.
7. Select 8, Config. Sat. Levels to configure saturation levels.
8. Repeat steps 3-6 to configure saturation levels.
AMS v7.0
Right click on the device, select "Device Configuration", then select
"Alarm/Saturation Levels," then "Alarm Levels" from the menu
1. Click Next after setting the control loop to manual.
2. Click Next after acknowledging the current alarm levels.
3. Select the desired alarm settings: NAMUR, Rosemount, Other
4. If "Other" is selected, enter desired "HI Value" and "LO Value" custom
values.
5. Click Next to acknowledge new alarm levels.
6. Select Next to acknowledge the loop can be returned to automatic
control.
7. Select Finish to acknowledge the method is complete.
8. Right click on the device, select "Device Configuration," then select
"Alarm/Saturation Levels," then "Saturation Levels" from the menu
9. Repeat steps 2 - 8 to configure saturation levels.
Transmitters set to burst mode handle saturation and alarm conditions
differently.
Alarm Conditions:
Analog output switches to alarm value
Primary variable is burst with a status bit set
Percent of range follows primary variable
Temperature is burst with a status bit set
Saturation:
Analog output switches to saturation value
Primary variable is burst normally
Temperature is burst normally
Rosemount 3051S Series
.
.
3-19

Advertisement

Table of Contents
loading

Table of Contents