Connecting

5.2 Unit with sensing head mounted LVDT conditioner card

5.2
Unit with sensing head mounted LVDT conditioner card
Note
Not applicable to Hazardous-rated units.
LVDT to SF500 Integrator
LVDT to LVDT conditioner card connections are made by Siemens.
*Where separation between the integrator and LVDT conditioner exceeds 150 m (500 ft):
1. Remove the jumpers from SF500 terminals 11/12 and 13/14.
2. Run additional conductors:
– from SF500 terminal 12 to conditioner terminal block marked "Integrator EXC+"
– from SF500 terminal 13 to conditioner terminal block marked "Integrator EXC-"
For further connection information on specific LVDTs, consult Siemens.
Shields are common, but not grounded to chassis. Run cable shields through SHLD terminals
and ground at Integrator only.
Note
Ensure that the connection between TB-2 and TB-17 is made.