In the course of system operation, technical and constructional
A . 1
changes may have been made to the MR system and its environ-
ment. It must be ensured that these components function satis-
factorily and do not present hazardous conditions.
After system acceptance by the customer, a daily visual inspec-
tion should be performed for the MR system with respect to the
following constructional changes:
Changes in the environment at the output of the exhaust
line (e.g. window installed retrospectively, inlet and outlet
of air-conditioning units, new buildings or temporarily
installed containers)
Changes to the air conditioning unit or venting system (e.g.
by adding air inlets and outlets in neighboring rooms)
Installation of additional MR systems (e.g. prohibited use of
the same exhaust line for several MR systems)
Constructional changes inside and outside the examination
room
The examination room must be checked for (newly introduced)
magnetic parts.
The annual technical safety inspections are listed in this system
A . 1
owner manual and may only be performed by Siemens Service.